分享

东芝数字胃肠(DSA)工作站ADR(NT)错误代码表

 zsok 2015-07-27

error code error description

S-1 Invalid state detected by ADR application S/W
S-2 Invalid state detected by NT operation system
S-3 Invalid state detected by Microsoft user interface S/W
S-4 I/O or internal resources operation time out expired
S-5 disk directory integrity loss detected while attempting to access image files
S-6 disk I/O error.physical disk sector read/write failure
H-100 no acknowledge from CCU
H-101 CMT use only
H-200 fail to complete PDB initialization
H-300 fail to complete RCB initialization
H-301 fail to complete RC-IO initialization
H-302 fail to load RCB look-up tables
H-303 communication problem between RCB and RC-IO
H-400 fail to complete keypad initialization
H-401 keypad communication failed
H-500 fail to complete LI initialization
H-600 EOD not responding to lock.unlock or eject requests
P-10 unable to access ADR parameter set
P-20 failed to access a given ADR parameter
P-30 wrong parameter type detected,parameters data integrity
P40 parameters set check sum error,parameters data integrity
P-50 invalid parameter relationship detected