Abstract

Purpose Acetabular component position is important for stability and wear. Fluoroscopy can improve the accuracy of acetabular component placement in the posterior approach and the direct anterior approach (DAA). The purpose of this study was to determine if the direct anterior approach in the supine position facilitates the accurate use of fluoroscopy and improves acetabular component position. 


Methods This retrospective, comparative study of 60 THAs with fluoroscopic guidance (30 in posterior approach group and 30 in DAA group) was performed by one surgeon from 2012to 2014 at a single institution.Demographic and perioperative data were compared using the Kolmogorov-Smirnov test to determine if they were statistically different.Thedifference between the measured intra-operative and postoperative values for both inclination and anteversion were analysed respectively. 


Results In the posterior approach group we found an average inclination on intra-operative fluoroscopy (IFluoro) of 36.8°± 3.72°, an average anteversion on intra-operative fluoroscopy (AFluoro) of 25.6°±3.64°, an average inclination on postoperativestandingAPpelvisX-ray(IAP X-ray)of39.29°±4.58°and an average anteversion on postoperative standing AP pelvis X-ray (AAP X-ray) of 21.31°±4.04°. In the DAA group we found an average DAA IFluoro of 42.32°±1.91°, an average DAA AFluoro of 22.3°±1.41°, an average DAA IAP X-ray of 42.98°±1.81° and an average DAA AAP X-ray of 22.88°± 1.38°. A difference was seen in variability using Kolmogorov-Smirnov test for inclination and anteversion with significant higher variation of measurements in the posterior approach group (p=0.022 and p<0.001 respectively). No statistically significant difference was seen in the DAA group using the fluoroscopy for inclination and anteversion. Conclusion Using fluoroscopy in the direct anterior approach, we achieved better intra-operative assessment of cup orientation resulting in decreased variability of acetabular cup anteversion than when used in the posterior approach.Atleast some of the improvement was due to the fact that the fluoroscopic image in the supine position was more accurate as measured against the postoperative standing AP pelvis. This study may influence the choice of approach in total hip replacement.


Keywords Directanteriorapproach(DAA) .Posterior approach .Fluoroscopy .Acetabularcomponent .Inclination . Anteversion


Introduction

A review of the literature on total hip arthroplasty would suggest that accurate placement of the acetabular component is both important and difficultto achieve. Methods are available to help ensure accurate component positioning including fluoroscopy[1,2],computer navigation[3,4], and robotics[5,6]. This study focuses on the use of fluoroscopy to help assure correct acetabular component orientation.


Multiple authors have assessed acetabular cup position in hip arthroplasty. Lewinnek et al. suggested a radiographic


definition for ideal acetabular component position as 15° (±10°) of anteversion and 40°±10° of abduction [7]. Cups placed outside this zone had a higher dislocation rate [8]. Furthermore, cups placed outside this zone also experience higher biomechanical stresses leading to increased rates of polyethylene wear and osteolysis [9, 10]. Hard-on-hard bearings are possibly more sensitive to acetabular malposition than hard-on-soft bearings [11]. With metal-on-metal implants, increased abduction and anteversion angles have repeatedly correlated with higher serum metal ion levels, a marker for wear [12–14].These studies support that both stability and wear are affected by cup position. 


Several studies conducted by experienced surgeons at prominent institutions have looked at the accuracy of acetabular component position when using traditional methods utilizing mechanical guides and anatomic landmarks such as the anterior superior iliac spine and pubic symphysis [7] and transverse acetabular ligament [15, 16]. In this setting, component position is dependent on the surgeon’s interpretation of the position of the patient’s pelvis on the operating table. Internal and external landmarks, which may be obscured [17, 18] by body habitus or the use of smaller incisions/minimally invasive techniques are also utilized [19]. 


A recent study by Barrack et al. [20] of 1,549 total hips found that only 88 % of acetabular components were within broadtarget range(abduction30–55°andanteversion5–35°). WhenCallananetal.[21]reviewed1,823hips,only38%met a more stringent component position target range (abduction 30–45°andanteversion5–25°).Thesestudies,andothers[22, 23], would suggest that techniques that improve acetabular component position could have value. 


A study by Rathod et al. [24] has shown that fluoroscopy with anterior hip arthroplasty reduced the variability of acetabular cup positioning compared with a non-guidedposterior approach. Another study by Beemer et al. [2] suggested that the use of fluoroscopy with posterior hip arthroplasty may increase accurate placement of acetabular components for surgeons performing amixofprimary,revision,andcomplicated total jointarthroplasties.From previousstudies,itisunclear if the value of fluoroscopy is similar when used in the posterior approach as compared to the anterior approach. It is the purpose of this study to determine if the direct anterior approach in the supineposition facilitates the accurate use offluoroscopy and improves acetabular component position.


Materials and methods

A prospectively maintained database of THAs performed by the senior surgeon (N. Stewart) at one centre from May 2012 to November 2014 was reviewed for this retrospective comparative study. A consecutive series of patients who underwent primary THAs, unilateral or staged bilateral, with a cementless hemispheric acetabular design and cementless tapered wedge femoral component (Stryker orthopaedics company) were included in the study. 


TheEMRwasusedtoobtaininformationfromeachpatient including laterality of the operatively treated hip; performing surgeon; age, sex, height, weight, and BMI of the patient; femoral head size utilized; acetabular cup outer diameter; surgical approach; and pre-operative diagnosis. Patients were required to have a standing postoperative digital anteroposterior pelvic radiograph of acceptable quality, according to criteria previously described by Callanan et al. [21]. Hips without adequate radiographs were excluded. This series consisted of 60 patients undergoing THA. The senior surgeon converted from a posterior approach to performing nearly all his primary hips from a direct anterior approach as of December26,2013. After this date, only patients with a severe Dorr type A femoral geometry were replaced with a posterior approach. We reviewed a consecutive number of patients with a direct anterior approach, of the first 34, 30 had adequately preserved intra-operative fluoroscopy films and a quality postoperative APpelvisX-ray and were included. Working back in time,we again neededto review 34 patients operated upon via the posterior approach to find 30 with adequately persevered intraoperative fluoroscopic films and quality postoperative study films. 


At posterior approach surgery,the patients were placed in a lateral decubitus position with firm supports on a peg board for the anterior pelvis, chest, sacrum and dorsal spine to keep the patient in a stable position throughout the surgery. Once the acetabulum component was placed, fluoroscopy was used to confirm placement. To ensure proper orientation of the fluoroscopic beam tangential to the pelvis,theC-arm was first adjusted to align the centre line of the sacrum with the pubic symphysis.Next,theC-arm was adjusted to make the shape of the obturator on the fluoroscopic image match the shape of the obturator on the pre-operative standingAPpelvisX-ray.Then the position of the acetabulum was assessed with afluoroscopic view that placed the cup near the centre of the screen while still allowing visualisation of the pubic symphysis to provide vertical reference of the pelvic position (Fig. 1). 


The anterior approach was performed using the Arch table extension (Orbiswiss company) on a slider table (Steris Corporation) surgical bed (Fig. 2). After anaesthesia, the patient was placed in a supine position. In the supine position, the midline of the sacrum was checked to see if it was aligned with the pubic symphysis.Then the projection of the obturator foramen was evaluated subjectively by the surgeon and the C arm manipulated to have the fluoroscopic image match the pre-operative standing AP radiograph view of the pelvis (Fig. 3a and b). At the time of acetabular insertion, a



Fig.1 Fluoroscopic image allows assessing the position of the acetabular cup in a posterior approach after final cup placement


fluoroscopy view with the cup near the centre, and the pubic symphysis in the view, was used to confirm cup placement (Fig. 4). With either approach, the target inclination was 40° (±10°) and anteversion was 15° (±10°). 


Geometric analysis was performed on both intra-operative fluoroscopic images of final cup position and six-week postoperative anteroposterior pelvic radiographs. Geometric


Fig.2 The ArchTable Extension is a special table attachment that allows controlled movement of the extremity during preparation of the femur and acetabulum.


Fig. 3 a Fluoroscopic image obtained with the patient in the supine position on the operating room table with the C-arm beam positioned such that the intraoperative image matches the preoperative image. b The shape of the obturator foramen and the pubic symphysis are noted by the surgeon.


assessmentof the radiographs was performed byanorthopaedic surgeon (WF. Ji) not involved in the surgeries. Measurements included cup abduction angle and cup anteversion angle. Pre and postoperative AP radiographs were taken with patients in the standing position with hips in neutral position, the radiation beam centred on the pubic symphysis, and filmfocus distance approximately 120 cm. 


A picture archiving and communications system (PACS) software (Fujifilm, Stanford, CT, USA) was used to make the measurements. With regard to validity and convenience for calculation, we chose the method described by Lewinnek and Widmer for measurement of anteversion of the acetabular component [25]. It has been shown to be a valid and reliable methodfor radiographic analysisofcup anteversion [26].Acetabular inclination on the postoperative film was measured as the angle between the interteardrop line and the long axis of the acetabular cup face. On the intra-operative fluoroscopic view, the pubic symphysis was used to determine the vertical orientation of the pelvis (Fig. 4).



Fig.4 Fluoroscopic image allows assessing the position of the acetabular cup in the direct anterior approach (DAA) during and after final cup placement.


The difference between the measured intra-operative and postoperative value for both inclination and anteversion were calculated. Variances (square of the SDs) were used to define the variability of the outcome measure. They were compared using the Kolmogorov-Smirnovtest to determine if they were statistically different. An independent t-test (two tailed) was used for normally distributed continuous data and the Mann–Whitney U test was used for nonparametric data. Chi-square and Fisher’s exact tests were used for comparing categorical data. A p value of 0.05 was set as the level of statistical significance. The 95 % limits of agreement method was used to assess the agreement between intra-operative fluoroscopy and postoperative standing AP pelvis. First, the mean and SD of the differences between the two methods were calculated. Second, the 95 % limits of agreement were calculated as mean difference ±1.96 SD. Plots of differences against means were used to examine the assumptions of the limits of agreement. Statistical analysis was performed using SPSS software (version 16; Chicago, IL, USA).

Results

Demographic and perioperative data are presented in Table 1. There was no significant difference in patient age, sex, operative side, diagnosis, cup size, or femoral head size between the two groups.

Mean acetabular inclination and anteversion values within the two groups are summarised in Table 2. A difference was seen in variability using Kolmogorov-Smirnov test for inclination and anteversion with significantly higher variation of measurements in the posterior approach group for both variables (p=0.022 and p<0.001 respectively). 


For the evaluation of both precision and bias of intraoperative fluoroscopy, the difference between the cup inclination measured by fluoroscopy(IFluoro) and cup inclination measured by postoperative standing APpelvisX-ray(IAP X-ray) was calculated. We did the same calculation between intraoperative anteversion(AFluoro) and postoperative anteversion(AAP X-ray). The mean difference and 95% confidence intervals for each of these measures are shown in Fig. 5a–d. 


The results in Fig. 5a and b demonstrate that 95 % of the time the fluoroscopic view in the supine position is between no more than 3.5° less or 2° more than the value found on a post operative standing AP X-ray. Similar values for anteversion are an underestimate of 3.8° to an overestimate of 2.2°. 


The results in Fig. 5c and d demonstrate that 95 % of the time in the posterior position the fluoroscopic measure of inclination isbetween 11.6°less and 6.0°morethanthe inclination on the standing AP X-ray. Similar values for anteversion are an underestimate of 4.0° and an overestimate of 11.0°.

Discussion

Acetabular component position is important for endoprosthesis survival [27, 28], wear [22], hip load [29], and range of motion (ROM) [30]. Multiple techniques have been promoted to enhance the accuracy of component positioning, including the use of fluoroscopy [1, 2, 24], computer navigation [31, 32], and robotic assistance [5, 6]. Of these three, fluoroscopy is the most widely available and least expensive to adopt. This paper compares the accuracy of




fluoroscopy to the postoperative standing AP radiograph in both the lateral and supine positions. 


Rathod et al. [24] demonstrated that an anterior approach with fluoroscopy allowed more accurate acetabular positioning than a posterior approach without fluoroscopy. It’s unclear if the improvement in that sense was due to the use of fluoroscopy or the change in approach. Beemer et al. [2] demonstrated that the use of fluoroscopy can improve component position in the posterior approach. Our results demonstrate that when using fluoroscopy in the anterior approach, we achieved better results than when used in the posterior approach (Fig. 6). At least some of the improvement was due to the fact that the fluoroscopic image in the supine position was more accurate as measured against the postoperative standing AP pelvis. Relying on the fluoroscopic guided direct anterior approach in the supine position, the DAA surgeons were able to improve their accuracy and consistency to position the acetabular component in the safe zone for inclination and anteversion. 


While our study shows that fluoroscopy is more accurate and consistent in the supine position, it’s unclear if the degree of improvement is clinically significant. While detailed longterm studies need to determine to what extent we must optimize component position, it’s hard to argue against a technique which improves accuracy if it is relatively safe and inexpensive. Compared to computer navigation and robotics, fluoroscopy is routinely available, and thus not expensive.


Fig. 5 Deviation of the intraoperative fluoroscopic values from postoperative standing AP pelvis values compared to their mean. a Direct anterior approach (DAA) inclination. b DAA anteversion. c Posterior approach inclination. d Posterior approach anteversion. Dotted lines represent the 95% limits of agreement (Bland-Altman graph)



Fig. 6 A six-week postoperative anteroposterior standing pelvic radiograph showing bilateral total hip arthroplasties (THAs). The right hip was placed by the same surgeon in the directanterior approach(DAA) using fluoroscopy assistance,whereas the leftwas placed in the posterior approach using fluoroscopy assistance. (Inclination: left 51.65, right 45.26; Anteversion: left 30.35, right 21.37)


The radiation dose, while not measured in this study, is generally quite small. 


This study may influence the choice of approach in total hip replacement. If a surgeon wishes to adopt a technology that assists with component positioning, and chooses fluoroscopy, using the direct anterior approach would make that technology more effective. There are other considerations in choosing an approach [21, 33–39], but these findings may factor into that discussion. 


Continued research into accurate component positioningis needed.Comparing fluoroscopic guidance to navigation,orto robotics, would help determine which technologies provide greater benefit. Even in skilled and experienced hands, accurate component positioning is not always achieved, and so technologies that help with this task will likely continue to gain acceptance.


Acknowledgments The authors wish to thank Carol for the fluoroscopic image and the standing AP pelvis X-ray acquisition, and Shen Jing M.D., for assisting the statistical analysis.


References

1. Matta JM, Shahrdar C, Ferguson T (2005) Single-incision anterior approach for total hip arthroplasty on an orthopaedic table. Clin Orthop Relat Res 441:115–124 


2. Beamer BS, Morgan JH, Barr C, Weaver MJ, Vrahas MS (2014) Does fluoroscopy improve acetabular component placement in total hip arthroplasty? Clin Orthop Relat Res 472:3953–3962. doi:10. 1007/s11999-014-3944-8


3. Ryan JA, Jamali AA, Bargar WL (2010) Accuracy of computer navigation for acetabular component placement in THA. Clin Orthop Relat Res 468:169–177. doi:10.1007/s11999-009-1003-7 


4. NoglerM,MayrE, KrismerM,ThalerM (2008) Reducedvariability in cup positioning: the direct anterior surgical approach using navigation. Acta Orthop 79:789–793. doi:10.1080/ 17453670810016867


5. NawabiDH,CondittMA,RanawatAS,DunbarNJ,JonesJ,Banks S, Padgett DE (2013) Haptically guided robotic technology in total hip arthroplasty: a cadaveric investigation. Proc Inst Mech Eng H 227:302–309 


6. Domb BG, El Bitar YF, Sadik AY, Stake CE, Botser IB (2014) Comparison of robotic-assisted and conventional acetabular cup placement in THA: a matched-pair controlled study. Clin Orthop Relat Res 472:329–336. doi:10.1007/s11999-013-3253-7 


7. Lewinnek GE, Lewis JL, Tarr R, Compere CL, Zimmerman JR (1978) Dislocations after total hip-replacement arthroplasties. J Bone Joint Surg Am 60:217–220


8. Jolles BM, Zangger P, Leyvraz PF (2002) Factors predisposing to dislocation after primary total hip arthroplasty: a multivariate analysis. J Arthroplasty 17:282–288 


9. Patil S, Bergula A, Chen PC, Colwell CW Jr, D’Lima DD (2003) Polyethylene wear and acetabular component orientation. J Bone Joint Surg Am 85-A(Suppl 4):56–63 


10. Little NJ, Busch CA, Gallagher JA, Rorabeck CH, Bourne RB (2009) Acetabular polyethylene wear and acetabular inclination and femoral offset. Clin Orthop Relat Res 467:2895–2900. doi: 10.1007/s11999-009-0845-3 



本文来自季卫锋的投稿

个人简介


医学博士、师承博士后,副教授,副主任中医师.中华中医药学会骨伤分会关节病委员、浙江省中西医结合学会骨质疏松分会委员。留学美国Chipppewa Valley、日本Juntendo university骨科及运动医学中心。


擅长:髋关节疾病、膝关节病、肩周炎、颈椎病、腰椎间盘突出症、骨质疏松相关疾病及运动损伤的诊治和中医药治疗。人工关节置换术、髋膝关节镜、腰椎间盘摘除术、创伤骨折内固定术、骨质疏松性椎体骨折的椎体成形术等。


主持和参与国家级、省部级等研究课题6项,获科研奖项多项,其中省政府科技进步奖3项.已在国际和国内学术期刊上发表论文20余篇,参编多部专著。


    本站是提供个人知识管理的网络存储空间,所有内容均由用户发布,不代表本站观点。请注意甄别内容中的联系方式、诱导购买等信息,谨防诈骗。如发现有害或侵权内容,请点击一键举报。
    转藏 分享 献花(0

    0条评论

    发表

    请遵守用户 评论公约

    类似文章 更多