illumina双色测序中碱基识别的弊端<QC Fail Sequencing » Illumina 2 colour chemistry can overcall high confidence G bases

 BIOINFO_J 2018-01-11

转载于https://sequencing./articles/illumina-2-colour-chemistry-can-overcall-high-confidence-g-bases/
<QC Fail Sequencing » Illumina 2 colour chemistry can overcall high confidence G bases>

说明:本文章阐述的是Illumina测序过程中存在的弊端,主要涉及到通过双色来区分碱基的测序系统。在basecalling过程中,双色系统会将无信号的情况视为碱基G,致使会出现multi-G的情况。同时此条件下G碱基往往具备较高的质量值,无法通过通常的过滤方法去除,可以通过cutadapt软件--nextseq-trim参数进行(http://cutadapt./en/stable/guide.html#)

Introduction

Illumina sequencing chemistry works by tagging a growing DNA strand with a labelled base to indicate the last nucleotide which was added.  4 different dyes are used for the 4 different bases and imaging the flowcell after each base addition allows the machine to read the most recently added base for each cluster.

In the standard version of this system used on HiSeq and MiSeq instruments each cycle of chemistry is followed by the acquisition of 4 images of the flowcell, using filters for each of the 4 emission wavelenths of the dyes used.  The time taken for this imaging is significant in the overall run time, so it would obviously be beneficial to reduce this.

With the introduction of the NextSeq system Illumina introduced a new imaging system which reduced the number of images per cycle from 4 to 2.  They did this by using filters which could allow the simultaneous measurement of 2 of the 4 dyes used.  By combining the data from these two images they could work out the appropriate base to call.

 

4 colour detection

Base G filter A filter T filter C filter
G Yes No No No
A No Yes No No
T No No Yes No
C No No No Yes

 

2 colour detection

Base A+C filter T+C filter
G No No
A Yes No
T No Yes
C Yes Yes

 

The problem here though is that there is a fifth base option, which is that there is no signal there to detect.  This could happen for a number of reasons:

  1. There is no priming site for this read, so no extension is happening.  If this happens in the first read then it just won’t detect a cluster, but in subsequent reads the cluster will be assumed to still be present.
  2. Enough of the cluster has degraded or stalled that the signal remaining is too faint to detect
  3. Something is physically blocking the imaging of the flowcell (air bubbles, dirt on the surface etc etc)

In the 4 colour system this extra option is easy to identify since you get no signal from anything, but in the 2 colour system we have a problem, since no signal in either channel is what you’d expect from a G base, so you can’t distinguish no signal from G.

What we find therefore on Illumina systems running this chemistry is that we see an over-calling of G bases in these cases, and this can cause problems in downstream processing of data.

The Symptoms

There are a couple of different ways in which this problem can manifest itself.  Possibly the most obvious is that you can get a huge over-representation of poly-G sequences in reads other than the first read (which is used for cluster detection).  From reports we’ve seen this seems to be most prevalent in the first barcode read, but could presumably affect any read where the initial priming of the read failed for any reason.

The example below is the top of a barcode splitting report for a sample which had a lot of barcodes.  You can see that the most frequently observed barcode for this sample had polyG for barcode1, and that this wasn’t expected.  You can also see that the second barcode was read correctly so the cluster itself was OK, but for some reason just didn’t prime for the barcode 1 read.

poly_g_barcodes

If we extract out the polyG sequences from the first barcode read and look at their quality distribution we see that the qualities are universally high, so there’s no way to distinguish these mis-calls by using the normal quality control mechanism.

polyg_qualities

 

This then leads to the second type of problem seen in these datasets.  This occurs when the signal from a cluster completely degrades.  In a 4 colour system this would result in a somewhat random base call, but a very low quality score to reflect the low level of signal.  In a 2 colour system the quality will initially fall as the signal degrades, but eventually the signal will effectively disappear and the sequencer will start calling high quality G bases.

The sequence below shows this effect:

@1:11101:2930:2211 1:N:0
ATTTATTATTAATTAAATATTAATAATAAATAGATCGGAAGAGCACACGTCTGAACTCCAGTCACTAGCTTAGCGCGTATGCCGTCGTCGGCGTGCAAAAAAAAAGGGGGGGGGGGGGGGGGGGGGGGGGGGGGGGGGGGGGGGGGGG
+AAAAAEEEEEEEEEEEE6EEEEEAEEEEEEEEEA/EE<EEEAEE/EAEEAEEEE6</EEEEEA/<//<///A/A//////</E<//////E///A/</A/<<A////A/E<EEEEEEEAEEE/EEEAEAEAEAE6/AEAEE<AAEAEE

It’s easier to see if you visualise the quality scores for this sequence

single_seq_quality

You can see that the quality degrades as expected initially, but then the signal disappears, either because of an extension of the general degradation or because some external factor blocked the imaging, and the sequencer starts to call G bases, the quality rises again.

The big issue this cases is that most trimming systems assume that quality loss is progressive from the 5′ end of a read to the 3′, so when deciding where to quality trim a read they start at the 3′ end and work back towards the 5′ until good quality sequence is detected.  In the types of read shown above the read will not be trimmed because the most 3′ sequence is high quality, so incorrect base calls will remain which will adversely affect the downstream analysis of this data.

 

Mitigation and Prevention

There are a few potential strategies for fixing this problem and some of the symptoms will be easier than others.  For unprimed reads which are completely polyG then these will probably not affect downstream analyses other than triggering QC alerts, so will have less of an impact once people are aware of their source.  For the progressive quality loss the need for a solution is more pressing.  Illumina already have a system in place which artificially downweights the quality scores for reads whose quality dips below a certain level.  It seems likely that they should be able to adapt their calling software to recognise reads where quality dips to be replaced with high quality poly-G and then artificially downweight the quality scores for the remainder of the read.  This same case can probably also be tacked by the authors of trimming programs where they might, for example, provide and option to exempt G calls from the assessment of quality and to trim 3′ Gs from reads as a matter of course.

Lessons Learnt

You can’t always trust the quality scores coming from sequencers, and if you see odd sequences in your libraries it’s worth investigating them as they may unearth a deeper problem.