Sources of error in blood gas analysis

Essentially, this chapter is a short list of reasons as to why your ABG (or venous biochemistry)  measurement may be wrong, i.e. not reflective of what is happening in the patient. In 99% of cases, it is a problem with the collection storage and transport of the sample, because these are factors which are subject to human input and thus human error. The self-calibrating blood gas analyser is a dutiful and dependable servant; some sort of failure in its internal workings will only rarely contribute to the error (and usually it will be because some idiot human has improperly calibrated it). Lastly, a tiny fraction of errors are begot by the physicochemical limitations of the measurement method (eg. when an absurdly high bromine concentration interferes with chloride measurement and returns a spuriously raised chloride concentration value). This whole thing enjoys a rich full discussion in the section dedicated to arterial blood gas analysis.

Sources of error due to inappropriate sample collection and handling

Source of error in blood gas collection and handling

  • Sample contaminated with bubbles

  • Published information on the errors of ABG sampling arising from the presence of air bubbles comes from Biswas et al (1982). After two minutes, the presence of air bubbles or froth in the syringe had resulted in a significant increase of PaO2 and a decrease in PaCO2.

  • Sample contaminated with venous blood

  • Sample clotted

  • Sample contains too much heparin (liquid heparin dilutes the sample, and causes pH changes)

  • Haemolysis en route to the ABG analyser

  • Inappropriate type of syringe used (gas-tight syringes are needed, rather than evacuated tubes)

  • The sample contains an excessive number of leukocytes, and they have consumed all the oxygen (eg. leukaemia blasts)

  • The sample took too long to transport, and blood cell metabolism has changed the gas concentration (see the section on delayed processing below)

  • The sample was transported to the lab by a pressurised pneumatic system - this has the tendency to amplify errors caused by gas bubbles (Woolley and Hickling, 2003).

  • Sample was chilled with ice (and then analysed at body temperature values)

  • Low temperatures also make the ABG syringe polymer more gas-permeable, allowing gases to exchange with the atmosphere.

Sources of error in venous biochemistry sample collection and handling

  • Clotted sample

  • Haemolysed by inappropriately small needle

  • Haemolysed by syringe vacuum

  • Inappropriate collection tube

The consequences of a delay in sample processing

Delayed venous sample processing

Question 3.2 from the second paper of 2015 presents the candidates with a venous blood sample which demonstrates characteristic features of a delay in processing. A good reference is a 2008 paper by Tanner et al, examining the delayed processing of samples collected in rural and remote areas. The studies have discovered that over 4-24 hours of storage various changes take place. These changes (and the reasons behind them) were as follows:

Escape of cellular contents due to haemolysis
  • Potassium increases. After 24 hours, Tanner et al found the original K  level of 3.8 increased to 8.0.

  • Phosphate increases. In the same study, the PO4- went from an average value of 1.36 to 4.36.

  • Total protein increases

  • LDH increases

Compartment shift 
  • Sodium decreases

Metabolism by live cells
  • Acidosis develops (bicarbonate drops)

  • LDH increases (also because of anaerobic metabolism)

  • Glucose decreases (consumed by RBCs)

  • Lactate increases (produced by RBCs)

Consequences of delay in arterial blood gas processing

An ideal reference for this exists in the Journal of Critical Care (Woolley and Hickling, 2003). It is unfortunately not available as free full text, but the author was able to get access to the article anyway, and summarise its contents.

  • Some of the changes listed above apply:

  • pH decreases

  • HCO3- decreases

  • SBE decreases

  • Lactate increases

  • Glucose decreases

  • PaO2 decreases and PaCOincreases as a consequence of RBC metabolism (in the experiments by Biswas et al (1982), the PaO2 fell by up to 40% in samples which were stored at room temperature for twenty minutes. This ameliorated by storing the sample at 4°C, but all sorts of other problems develop as a consequence of this).

The actual difference in reality is not that great.  It is possible to demonstrate this directly, in the following scenario. The two gases are arterial, taken at 35% FiO2 from an ICU patient; the sample was analysed once, then left next to the ABG machine for a half-hour or so, and then accidentally analysed again by someone who thought it was fresh.

The sample on the left was first, the one on the right was the thirty minute delay. The ambient temperature was probably close to 20 degrees. As one can see, the differences are trivial. During that half hour, this anaemic patient's few remaining RBCs managed to metabolise only 0.1mmol/L of glucose, and produced a barely measurable increase in lactate and CO2.

Instrument errors:

  • Inappropriate electrolyte temperature

  • Poorly calibrated ABG analyser

  • Inadequate quality control and maintenance of the electrodes (they have a measurable lifespan; for instance the Clark electrode needs to be changed every 3 years as its Teflon membrane becomes coated with proteinaceous filth).

  • Poorly compensated Clark electrode nonlinearity.

Limitations of measurement technique

pH measurement

Heparin of any sort: Given its extreme anionic charge and near-total dissociation at physiologic pH, it would be surprising if heparin failed to interfere in some way with the strong ion difference in a blood gas sample. Indeed, it would appear that even a 'thin film' of it skews the results in the direction of a metabolic acidosis. The quoted study is from 1985, a grim and primitive age when intensive care staff would have to manually heparinise the ABG syringe prior to collecting the sample, sending it to a distant analyser with a porter.

Sodium measurement

Lithium heparin: Cations other than sodium could potentially exist in the sample in such concentrations as to fool the supposedly sodium-selective ceramic electrode membrane. Of these the most likely culprit is lithium from lithium heparin. These days the arterial blood gas syringe typically contains a very small amount of dry heparin, about 20-25 units per ml.

Chloride measurement

Halogen ions: the measured Cl- increases in presence of halogen ions, such as bromine and iodine. This is a selectivity failure of the supposedly selective chloride-sensitive electrode membrane, which will accept any halogen if their concentration is high enough. 

Salicylates: in massive salicylate overdose there is enough salicylic acid around to compete with chloride for the ionophore in the chloride-selective electrode membrane.

Haematocrit measurement

Haemolysis: because the ABG machine shreds the red blood cells with 30Hz ultrasound in order to get to their haemoglobin, it will not notice if there is already free haemoglobin in the bloodstream (eg. due to haemolysis).  

Carboxyhaemoglobin measurement

 Foetal haemoglobin and carboxyhaemoglobin have a very similar absorptivity spectrum. Thus, foetal haemoglobin interferes with the spectrophotometric measurement of carboxyhaemoglobin and falsely increases the measured FCOHb. Conversely, the presence of carbon monoxide poisoning in a pregnant woman might mask a foeto-maternal haemorrhage.

Lactate measurement

The amperometric measurement of lactate using the lactate-sensitive electrode relies on the use of lactate oxidase. This enzyme catalyses the reaction which converts lactate into pyruvate, producing hydrogen peroxide which is reduced at the measurement cathode. Glycolic acid, the metabolic byproduct of ethylene glycol metabolism, also acts as the substrate for this enzyme. Therefore, in ethylene glycol poisoning the lactate measurement by the blood gas analyser will be spuriously elevated. The formal lactate measurment by use of a lactate dehydrogenase enzyme assay will still yield a correct result. The difference between the 'formal' and the ABG lactate is described as the 'lactate gap', and is a well known phenomenon of ethylene glycol toxicity. This source of ABG error is made famous by its use in a classic ABG question (Question 1.12) from the 2003 edition of Data Interpretation in Intensive Care Medicine by Bala Venkatesh et al. It had also made an appearance in Question 20.2 from the second paper of 2017. Interestingly, the Reference Manual for the local ABG analyser lists a large number of molecules which can potentialy cause interference with lactate measurement- notably ascorbic acid, bilirubin, citrate, EDTA, ethanol, heparin, glucose, paracetamol, salicylate and urea.

Methaemoglobin measurement

Hyperlipidaemia: which in general tends to produce unusual reading in the spectrophotometer, giving rise to abnormally high total haemoglobin readings and negative carboxyhaemoglobin numbers. These extra fatty acids tend to scatter the light, interfering with the absorption spectrum and giving the impression that a lot more light has been absorbed. This can be seen in central venous gases taken accidentally from the TPN lumen. The elevation is typically modest - the author's own recent experience with a pancreatitis patient yielded a serum triglyceride level of 39.9, and a paltry FMetHb of 2.4%. One would obviously need a truly monstrous hyperlipidaemia.
Isosulfant blue: otherwise known as 'patent blue' - a dye used in surgery to identify lymphatic vessels, eg. for sentinel node biopsy- causes the same sort of problem: its blue enough to interfere with spectrophotometry.